Draft Technical Specifications for the Procurement of Volumetric Infusion Pumps under the Centralized Public Procurement for Medical Equipment Orders using Price-Quality Ratio Criteria via the CPO LT Electronic Catalogue AI

TEDAS your procurement assistant

Ask a question about this procurement

t

This information is intended for suppliers interested in the draft technical specifications for the purchase of volumetric infusion pumps via the CPO LT electronic catalogue.

Subject of Procurement

  • Items purchased: Volumetric infusion pumps (CPV code 33100000).
  • Key technical requirements:
    • Operating modes: continuous infusion and transfusion.
    • Flow rate settings: 0.1–1200 ml/h, accuracy tolerance ± 5%.
    • Occlusion pressure limits: 0.1–1 bar (or 250–900 mmHg).
    • Battery life: at least 10 hours…
       