Here are the current procurement conditions, prepared according to the submitted documents: # Procurement Object * What is being purchased: Neutral disposable electrodes (indicative quantity – 3000 units). * Main technical requirements: * Must be technically compatible with BOWA monopolar electrosurgical devices used in the hospital (manufacturer’s confirmation required). * Disposable, universal (for patients ≥ 5 kg), split (dual-contact surface), anatomical shape. * Conductive area: 110 cm² ± 5%. * Ultra-thin water-based gel, liquid-resistant coating,…