Here is the most important information for the supplier regarding the procurement:# Object of procurement* Purchased: Dental equipment – portable dental X-ray unit with digital X-ray (visiography) system.* Key technical requirements:* Focal spot size of the radiation source: ≤ 0.5 mm.* Exposure time: 0.05–0.5 s.* Tube voltage: ≥ 60 kV; Anodic current: ≥ 2 mA.* Package must include a protective collar (lead equivalent ≥ 0.35 mm Pb).* Weight of X-ray unit: ≤ 2 kg.* Digital…