Here is the key information for the supplier regarding the ongoing market consultation for the purchase of a pulmonary diagnostic system:
Subject of Purchase
What is being purchased: Pulmonary diagnostic system (goods).
Key technical requirements:
The system must perform spirometry (slow and forced with bronchodilation tests), lung volume (FRC, RV, TLC), and diffusion capacity (DLCO, KCO) measurements.
Flow sensor: ultrasonic or equivalent (accuracy ±3%, range 0-18 l/s).