To be procured: Medical equipment – Electrosurgical generator (2 units).
Main technical requirements:
Monopolar cutting (Pure/Blend): Pure cut (without coagulation effect, not less than 100%); Blend cut (with coagulation effect, not less than 50%); High coagulation (with strong coagulation effect, not less than 25%).
Maximum power/impedance (Monopolar): Pure – not less than 300 W at 300 Ω; Blend – not less than 200 W at 300 Ω;…