Procured item: Goods (Medical devices for neurosurgery).
Main technical requirements, peculiarities:
Lumbar drainage system: Radiopaque lumbar catheter (OD 1.5 – 1.6 mm, length $\ge 80$ cm), Tuohy type 14G needle with Huber tip, Luer Lock connection, drainage bag ($\ge 600$ ml), extension line ($\ge 165$ cm).
External drainage system with ventricular catheter: Drip chamber ($\ge 100$ ml) with antireflux valve, pressure scale (cm $\text{H}_2\text{O}$ and mm $\text{Hg}$),…